Fisher-paykel MR850 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Płyty grzewcze Fisher-paykel MR850. Fisher & Paykel MR850 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - HUMIDIFIER

MR850 RESPIRATORY HUMIDIFIER Technical Manual REVISION J Copyright ©2005 Fisher & Paykel Healthcare Ltd. Auckland, New Zealand

Strona 2

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 10 3 Specifications 3.1 Mechanical Dimens

Strona 3 - TABLE OF CONTENTS

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 11Non-heater wire operation: Invasive Mod

Strona 4

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 12 4 Operating Modes and Controls 4.1 Hu

Strona 5

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 134.1.2 Non-Heater Wire Operation (Softw

Strona 6

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 14 Stand-by (Software versions: 5.45 &

Strona 7 - Introduction

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 15then the chamber set point will automat

Strona 8 - 1.4 Product Application

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 16 connected. Breathing circuit recognit

Strona 9 - 2 Humidifier Symbols

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 174.2.3 Mute Button The mute button sile

Strona 10 - 3 Specifications

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 18 4.4.2 Temperature / Flow Probe Connec

Strona 11 - 3.6 Standards and Approvals

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 194.4.5 Water Out Indicator This indica

Strona 12 - 4.1 Humidifier Operation

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 2 Fisher & Paykel Healthcare Head Of

Strona 13

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 20 Figure 4.3 Temperature vs Time to Ala

Strona 14

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 215 Maintenance Procedures In order to ke

Strona 15

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 22 5.2 Safety Check The unit should be t

Strona 16 - 4.2 Humidifier Controls

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 23 5.3 Cleaning Instructions 5.3.1 MR850

Strona 17 - 4.4 Setup Indicators

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 24 6 Troubleshooting 6.1 Operational Prob

Strona 18

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 25Symptom Corrective Action Reference H

Strona 19 - 4.5 Operational Alarms

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 26 Symptom Corrective Action Reference

Strona 20

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 276.3 “See Manual” Error Codes The follo

Strona 21 - 5 Maintenance Procedures

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 28 Error Description of Fault E41 Faile

Strona 22 - 5.2 Safety Check

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 29 6.4 Diagnostic Menu By pressing the mu

Strona 23 - 5.3 Cleaning Instructions

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 3 TABLE OF CONTENTS 1 INTRODUCTION ...

Strona 24 - 6 Troubleshooting

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 30 6.4.2 Diagnostic Menu for Software Ve

Strona 25

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 316.4.3 Diagnostic Menu for Software Ver

Strona 26 - 6.2 Technical Problems

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 32 6.4.4 Diagnostic Menu for Software Ve

Strona 27

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 336.4.5 Diagnostic Menu for Software Ver

Strona 28

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 34 6.4.6 Diagnostic Menu for Software V

Strona 29 - 6.4 Diagnostic Menu

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 357 Servicing Procedures 7.1 General Cons

Strona 30 - Display Description

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 36 Figure 7.2 Showing PCB fasteners 7.2.

Strona 31

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 37MR850 Model Number Supply Voltage Fus

Strona 32 - Display Description

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 38 Figure 7.4 Showing Humidifier Power P

Strona 33

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 393. If the thermal cutout "clicks&

Strona 34

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 4 6 TROUBLESHOOTING ...

Strona 35 - 7 Servicing Procedures

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 40 Figure 7.7 Showing location of Heater

Strona 36

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 415. Remove the heater element, leaving

Strona 37

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 42 7.2.6 Installing New Software NOTE: A

Strona 38

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 436. Apply mains power to the humidifier

Strona 39

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 44 8 Performance Testing This section dis

Strona 40

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 451. Hold down the power button, then ap

Strona 41

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 46 8.1.4 Humidifier Voltage Calibration

Strona 42

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 473. Connect the humidifier chamber inle

Strona 43

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 48 1. Perform a humidifier calibration

Strona 44 - 8 Performance Testing

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 499 Recommended Maintenance Checklist Th

Strona 45

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 5 Revision Description of Technical Manu

Strona 46

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 50 (Blank)

Strona 47 - 8.2 Probe Accuracy Check

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 5110 Spare Parts Should any parts of the

Strona 48

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 52

Strona 49

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 53 Heater Plate Assembly Item Part Num

Strona 50 - (Blank)

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 54

Strona 51 - 10 Spare Parts

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 5511 Calibration Probe The information pr

Strona 52

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 56 12 Serial Port & Logging Softwar

Strona 53 - Heater Plate Assembly

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 5712.3.2 Logging Humidifier Data to File

Strona 54

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 58 13 EMC INFORMATION WARNING: The use of

Strona 55 - 11 Calibration Probe

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 59 Guidance and manufacturer’s declaratio

Strona 56 - 12.3 OPERATING INSTRUCTIONS

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 6 Revision Description of Technical Manu

Strona 57

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 60 14 Product Change History Due to upg

Strona 58 - 13 EMC INFORMATION

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 61 History Change for model GJU PCB Versi

Strona 59

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 62 History Change for all model except JH

Strona 60 - 14 Product Change History

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 7Introduction 1.1 About this Manual This

Strona 61

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 8 Expiratory Limb The section of the bre

Strona 62

MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal) 92 Humidifier Symbols Caution: Hot surfa